Documentation

How the simulation behaves

What each value you set does, when the monitor shows something else, and where the simulation departs from real patients and monitors.

What the patient and monitor do follows real physiology and bedside monitors unless marked Simplified, where the simulation deliberately departs from real patients or monitors for teaching or ease of use. In reality then says what actually happens.

Principles

Values are independentSimplified
Changing one value never changes another: lowering SpO₂ does not raise HR, and stopping breathing does not lower SpO₂. You drive every change. The only links are the rules on this page.

In reality: Vital signs drive each other: hypoxia and hypotension raise the heart rate, apnea lowers SpO₂, and so on.

Waveforms follow the heartbeat
The pleth, Arterial line and CVP traces are timed against each QRS, and the CVP’s waves change with the rhythm (see CVP).
Set points are kept
When a rule overrides a reading (for example, no pulse), your value is kept and returns when the rule no longer applies. A Display: … line under the field on the Remote says what learners see instead, and why.
Waveforms are illustrativeSimplified
Traces are plausible teaching shapes, not diagnostic recordings.

In reality: Waveforms differ from patient to patient, and with lead placement, transducer setup and artifact.

Disconnected devicesSimplified
A disconnected device shows no trace or reading. Its values are kept for when it is reconnected.

In reality: Monitors show a lead-off or no-sensor message and alarm.

Heart rate and rhythm

HR counts QRS complexes
HR is the ventricular rate. For irregular rhythms (AF, MAT, ectopy, Mobitz) the reading is an average that wanders a little, as a monitor’s does.
No rate to countSimplified
VF shows HR as —. Asystole shows HR 0.

In reality: Monitors may count VF undulations or artifact as a rate; the alarm, not the number, flags VF.

Atrial activity runs on its own
The P waves of complete heart block ignore the ventricular rate, and continue without it. So do flutter waves when there are no QRS complexes.
FlutterSimplified
Flutter waves run near 400/min (infant), 300/min (child), 300/min (adult), adjusted slightly so each QRS follows a whole number of them (2:1, 3:1, 4:1…). If no whole ratio fits your HR, the ratio alternates so the average matches it.

In reality: The flutter rate (about 250–350/min in adults) does not adjust to the ventricular rate.

Complete heart block atrial rateSimplified
P waves run at 140/min (infant), 100/min (child), 75/min (adult).

In reality: The atrial rate varies like any sinus rate.

PR by age
Shorter in children, and slightly shorter at fast rates. Sinus beats share one normal PR, with or without ectopy; AV block PRs stay longer.
Fixed AV block patternsSimplified
First-degree block has a PR of 0.32 s in adults. Mobitz I and Mobitz II always conduct 4:3.

In reality: PR prolongation varies, and block ratios (3:2, 4:3, 5:4…) vary and can change from moment to moment.

AV block P waves stay visibleSimplified
At faster rates the PR shortens just enough to keep every P wave, including blocked ones, clear of the T wave before it, while staying long (first-degree) or visibly lengthening (Mobitz I). Only near the top of each rhythm’s range does a P wave ride on the T wave.

In reality: At fast rates P waves often hide in the T wave before them, which is part of what makes AV block hard to spot.

Pulse deficit
A beat that ejects too little gives the oximeter no pulse: no pleth beat and no pulse tone. Pulse counts only the beats it detects, so it reads below HR. The Arterial line still shows a small pressure beat.
Which beats go undetectedSimplified
Every PVC is missed, so bigeminy reads Pulse at half the HR and trigeminy at two thirds. In AF, a beat after an interval well below the average misses too, once the beat is short enough: none up to about 130/min in adults, then about a fifth of beats from 160/min (only at higher rates in children and infants). Every other rhythm detects every beat.

In reality: Whether a weak beat registers varies from beat to beat, with its timing, the patient’s filling and the oximeter.

Weaker early beats
A beat after a short interval fills less, and a PVC ejects less, so their pulse beats are smaller.
Rhythm changes are immediate
The onset or conversion of a rhythm is abrupt, so a rhythm change and its HR always apply at once, even with gradual Timing.
Choosing a rhythm
Stages that rhythm’s typical HR for the patient’s age, as an editable pending change. Sinus with ectopy or first-degree block keeps your HR when it fits.
A new rhythm restores the pulseSimplified
Choosing another rhythm turns the pulse back on, so a shock to sinus rhythm is ROSC.

In reality: An organized rhythm after a shock can still be PEA. Set Patient has a pulse? to No again for that.

The Display never names the rhythmSimplified
Learners read it from the ECG.

In reality: Many monitors name arrhythmias in their alarm messages.

No pulse

When there is no pulse
VF or asystole, or an organized rhythm with Patient has a pulse? set to No (PEA, pulseless VT). HR 0 counts too.
What the monitor can measure
  • ECG keeps drawing the rhythm; HR still counts complexes, so PEA shows its rate.
  • SpO₂ and Pulse: no signal, flat pleth.
  • Pulse tone: silent.
  • NIBP: a completed cycle gives no reading.
Arterial line reads 0Simplified
The Arterial line reads 0/0 (0) on a flat trace.

In reality: Without cardiac output, arterial and venous pressures settle toward a shared mean filling pressure, typically about 10–20 mmHg, not 0.

CVP keeps the set valueSimplified
CVP keeps reading your value, on a flat trace.

In reality: CVP drifts toward the same shared filling pressure.

RR, EtCO₂ and temperatureSimplified
The no-pulse rule leaves them to you. The arrest presets below stage the usual changes.

In reality: Without circulation, breathing stops within moments (agonal gasps aside) and EtCO₂ falls toward 0.

Arrest presets
Losing the pulse stages the values in the table below on the Remote as pending changes. Edit them before Apply if needed (for example, ventilations during CPR).
When the pulse returns
The Remote stages the pre-arrest RR, EtCO₂ and Arterial line values for any that still hold an arrest or CPR default. Values you set during the arrest stay. Turn off Restore pre-arrest values when the pulse returns in Monitor settings to keep current values instead.
Arrest presets
Every age
RR (/min)0
EtCO₂ (mmHg)0
Arterial line (mmHg)0/0

CPR

Starting CPR
Start CPR acts immediately, with or without a pulse. Pause CPR stops compressions for rhythm checks.
Losing the pulse during CPR
If the pulse is lost while compressions run, the Remote stages the values in the table instead of the arrest zeros, so the Arterial line, ventilations and EtCO₂ continue.
Compression artifact
Compressions add large, uneven artifact over the underlying rhythm on the ECG.
HR during CPRSimplified
HR shows —.

In reality: Monitors often count compressions as the heart rate; the number means nothing during CPR.

Arterial line
Each compression pushes the Arterial line from the set diastolic to the set systolic at the Compression rate, even without a pulse.
SpO₂ and pulse toneSimplified
Still no signal, as without a pulse.

In reality: Some oximeters pick up compressions and show a pleth and a number that mean little.

CVPSimplified
Stays flat at your value.

In reality: Each compression also drives a large right atrial pressure wave.

Starting valuesSimplified
Starting CPR replaces arrest zeros with values for the patient’s age (below), so compressions show something to coach from: arterial diastolic pressures around general targets for effective compressions (these vary by guideline and patient), and ventilation rates for an advanced airway. Values you already set are kept.

In reality: Pressures and EtCO₂ during CPR depend on compression quality and often fall short of these targets.

CPR qualitySimplified
Arterial line, EtCO₂, RR and Compression rate are your levers; they need Apply like any edit.

In reality: Pressures and EtCO₂ change compression by compression with depth, rate, recoil and rescuer fatigue.

ROSCSimplified
An Apply that restores the pulse stops CPR.

In reality: Compressions stop when a rhythm check finds a pulse.

Values Start CPR fills in
InfantChildAdult
Arterial line (mmHg)55/2565/3060/20
RR (/min)302510
EtCO₂ (mmHg)151515

Breathing

Apnea
RR 0 means no breaths: a flat capnogram and EtCO₂ reading 0. Your EtCO₂ value returns with breathing.
The last breathSimplified
When RR goes to 0, the breath in progress finishes its exhalation, then the capnogram falls to baseline without the pause it would have had: the trace is flat within 4 s at most, at any RR. The EtCO₂ reading shows 0 at once.

In reality: Without a next breath to wash it out, exhaled gas can linger at the sensor and the capnogram fall slowly.

One respiratory rate
The respiration and capnography waveforms both follow RR.

Blood pressure

Two separate devices
NIBP (cuff) and the Arterial line measure separately.
Keep BP correlatedSimplified
With the Arterial line and cuff both connected, NIBP follows the Arterial line at a fixed percentage offset (default 0%) and is read-only; edit the Arterial line instead. If the offset would push NIBP past its limits, it flips sign. Turning the link off keeps both sets of values.

In reality: Cuff and arterial readings differ by varying amounts; cuffs tend to read high at low pressures and low at high pressures.

MAPSimplified
Calculated as (systolic + 2 × diastolic) ÷ 3, rounded, unless you enter a MAP yourself.

In reality: Cuffs measure MAP directly and estimate systolic and diastolic from it; arterial monitors average the waveform. (S + 2D) ÷ 3 is a bedside estimate.

The cuff measures only when cycled
The NIBP reading updates only when a cycle finishes (Cycle NIBP, or Update Now to finish early) and is held until the next. Applying new BP values does not update it.
Automatic cycling
Off by default. With Automatic cycling set in Monitor settings (every 1, 2, 3, 5, 10, 15 min), the cuff starts a cycle by itself one interval after the last cycle started, manual or automatic. Changing the interval or reconnecting the cuff restarts the countdown. The Display shows the interval on the NIBP reading.
No NIBP reading
Until the first completed cycle, after the cuff is disconnected, and after a cycle with no pulse.

CVP

Your value, the rhythm’s waveformSimplified
The CVP value changes only when you change it. Its waveform follows the ECG rhythm and the pulse (below).

In reality: CVP moves with volume status, heart function, breathing and ventilation.

The reading
A single mean pressure in mmHg, as monitors show. Venous pressure needs no pulse.
Cannon a waves
When the atria contract against a closed tricuspid valve (during ventricular systole), the a wave is a tall cannon wave: on every beat in junctional rhythm, SVT and ventricular pacing, where the atria are activated backwards; on each PVC; and now and then in complete heart block and ventricular rhythms, whose atria beat on their own.
Wave sizeSimplified
The waves are the same size at any CVP and balance so the trace averages your value. They shrink below 5 mmHg so the trace never reads below 0.

In reality: Wave size varies with the heart and the line, and a spontaneous breath can pull CVP below 0.

Not currently simulatedSimplified
Large v waves (tricuspid regurgitation), tamponade and constriction patterns, breathing variation, flutter or fibrillation waves in the trace, and compression waves during CPR.

SpO₂ and pulse tone

Pleth dampeningSimplified
One control shrinks and rounds the pleth to suggest poor perfusion. At 100% the oximeter loses the signal and SpO₂ shows no reading.

In reality: The pleth does shrink and round with poor perfusion, from many causes (cold, vasoconstriction, low output), and oximeters lose the signal at different points.

Pulse tone
One tone per detected beat, its pitch falling as SpO₂ falls. It needs a connected pulse oximeter, a pulse and a pleth signal.
Pulse tone pitchSimplified
Pitch falls by an equal ratio per percentage point from the 100% tone (default 880 Hz) to the 70% tone (500 Hz), and stays at the 70% tone below that. Set both in Monitor settings.

In reality: Each manufacturer uses its own pitch scale.

Natural variability

Readings fluctuate
Vital signs drift a little from moment to moment, so the simulation adds small, slow fluctuations around your set points. On by default at strength 50, under Simulation options.
What variesSimplified
Each on its own: HR (sinus rhythm, with or without ectopy or first-degree block; other rhythms keep their rate), SpO₂, RR, EtCO₂ and Arterial line. NIBP, CVP and temperature never vary.

In reality: Fluctuations are linked: HR rises with inspiration, and arterial pressure and CVP vary with breathing.

How muchSimplified
About the amounts below at strength 50; 100 doubles them. Your set points never move.

In reality: How much vitals fluctuate differs between patients, and falls with illness and sedation.

Same everywhere
Every Display shows the same fluctuations.
Typical fluctuation at strength 50
About ±
HR2 bpm
SpO₂1%
RR1/min
EtCO₂2 mmHg
Arterial line3/2 mmHg

Gradual changes and readings

Gradual ApplySimplified
With 5–60 s under Timing, each changed vital moves in a straight line to its new value on its own clock. A later Apply re-times only the vitals it changes.

In reality: Real changes rarely move in straight lines or at a steady rate.

Always immediate
Rhythm and HR together, the pulse setting, device connections, CPR, and audio.
Losing the pulse
An arrest is no slow decline. The Apply that loses the pulse changes RR at once (the breath in progress finishes its exhalation, with no new ones at RR 0) and brings EtCO₂ down within 10 s, whatever the Timing. Later Applies during the arrest use the Timing as usual.
Readings update like a monitor
Monitors average and refresh their numbers. With Sample and hold numerics (Monitor settings), HR, Pulse, Arterial line and CVP update about once a second; SpO₂, RR and EtCO₂ about every two seconds.

Patient age

Normal vitals
Typical values within the AHA PALS and ACLS ranges for each age (below).
Choosing an age
Starts a new patient: it stages that age’s normal vitals, sinus rhythm with a pulse, and the default devices as pending changes. Nothing of the previous patient carries over; only Simulation options and audio stay.
Age tunes suggestions
The HR staged when picking a rhythm, and the values Start CPR fills in. Age never links one vital to another.
A new age is a new patient
Applying it is immediate, whatever the Timing: CPR stops, every Display clears its traces and the NIBP reading, and the automatic NIBP countdown restarts.
Normal vitals by age
InfantChildAdult
HR (bpm)14010080
RR (/min)382216
SpO₂ (%)989898
EtCO₂ (mmHg)383835
BP (mmHg)80/50100/62120/80
CVP (mmHg)556
Temperature (°C)373737